Z-0589-2011

GE Responder 2000 Defibrillator/Monitor, models 2025653, 2026109, and 2026114 that were manufactured between...

ZOLL Medical Corporation / initiated 2010-04-28 / Terminated / root cause: manufacturing process

Cardiac Science Corporation began a recall of GE Responder 2000 Defibrillator/Monitor, models 2025653, 2026109, and 2026114 that were manufactured between July 25, 2008 and October 2, 2009 by GE Healthcare on . The reason FDA records is "Some units may be non-operational during a rescue, due to a defective PCB interconnect on traces connected to the reed relay. This will result in an inability to analyze the patient's heart rhythm or deliver therapy...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0589-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZOLL Medical Corporation
Product
GE Responder 2000 Defibrillator/Monitor, models 2025653, 2026109, and 2026114 that were manufactured between July 25, 2008 and October 2, 2009 by GE Healthcare.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Some units may be non-operational during a rescue, due to a defective PCB interconnect on traces connected to the reed relay. This will result in an inability to analyze the patient's heart rhythm or deliver therapy during a rescue.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Cardiac Science issued a Urgent Medical Device Corrective Action letter dated April 28, 2010 requesting GE Healthcare cease distribution of affected devices and to instruct end users of distributed devices to schedule a service call to have the device corrected. Cardiac will send rework kits no later than June 30, 2010. CSC can be contacted at 44 161 926 0011 to arrange return material authorization and replacement of affected devices.
Quantity in commerce
130 units
Distribution
Worldwide Distribution: Throughout the US, and in the countries of Germany, Korea, Peru, Singapore, and Spain.
Product codes and lots
Serial Numbers:  3201281, 3201806, 3201810, 3201815, 3202223, 3202233, 3202244, 3202248, 3202249, 3202253, 3202255, 3202256, 3202257, 3202260, 3202262, 3202269, 3202271, 3202273, 3202275, 3202280, 3202281, 3202283, 3202284, 3202285, 3202286, 3202287, 3202288, 3202290, 3202291, 3202292, 3202297, 3202299, 3202300, 3202301, 3202303, 3202304, 3202305, 3202306, 3202308, 3202310, 3202311, 3202312, 3202313, 3202315, 3202316, 3202317, 3202319, 3202321, 3202324, 3202327, 3202328, 3202331, 3202332, 3202333, 3202334, 3202336, 3202337, 3202338, 3202339, 3202340, 3202344, 3202347, 3202348, 3202349, 3202350, 3202351, 3202352, 3202354, 3202355, 3202356, 3202357, 3202358, 3202360, 3202361, 3202362, 3202363, 3202364, 3202365, 3202371, 3202377, 3202379, 3202380, 3202381, 3202382, 3202385, 3202386, 3202387, 3202388, 3202389, 3202390, 3202391, 3202392, 3202393, 3202394, 3202395, 3202396, 3202397, 3202398, 3202399, 3202400, 3202401, 3202403, 3202404, 3202405, 3202406, 3202407, 3202408, 3202409, 3202411, 3202412, 3202414, 3202415, 3202416, 3202417, 3202418, 3202419, 3202420, 3202421, 3202422, 3202423, 3202427, 3202429, 3202439, 3202442, 3202443, 3202444, 3202447, 3202448, 3202450, and 3202451.