Z-0924-2007
Terumo Advanced Perfusion System 1, 6 inch diameter Roller Pump; Model 801041
Terumo Cardiovascular Systems Corporation / initiated 2004-04-01 / Terminated
Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1, 6 inch diameter Roller Pump; Model 801041 on . The reason FDA records is "A belt slip message may be displayed on the roller pump due to low belt tension or the presence of grease". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0924-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Advanced Perfusion System 1, 6 inch diameter Roller Pump; Model 801041.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- A belt slip message may be displayed on the roller pump due to low belt tension or the presence of grease.
- FDA root cause
- Other
- Action
- The firm is visiting consignees and making corrections as complaints are received. These corrections, to date, have been made between 04/04 and 09/06. The firm will correct other units as complaints are received and issued a Device Correction Letter dated 6/6 /07 notifying all users of the problem and appropriate responses should it occur.
- Quantity in commerce
- 525 total for both models
- Distribution
- Worldwide: USA, Belgium, Canada, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Korea, Malaysia, Singapore and United Arab Emirates.
- Product codes and lots
- Serial numbers 0033 through 0355 are potentially affected by the problem. Serial numbers 0033, 0041, 0049, 0052, 0105, 0115, 0117, 0119, 0122, 0145, 0151, 0159, 0160, 0161, 0162, 0170, 0174, 0218, 0226, 0242, 0257, 0262, 0277, 0279, 0290, 0299, 0300, 0305, 0306, 0331, 0334, 0339, 0343 and 0360 have been corrected.