Z-0923-2007

Terumo Advanced Perfusion System 1, 4 inch diameter Roller Pump; Model 801040

Terumo Cardiovascular Systems Corporation / initiated 2004-04-01 / Terminated

Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1, 4 inch diameter Roller Pump; Model 801040 on . The reason FDA records is "A belt slip message may be displayed on the roller pump due to low belt tension or the presence of grease". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0923-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1, 4 inch diameter Roller Pump; Model 801040.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
A belt slip message may be displayed on the roller pump due to low belt tension or the presence of grease.
FDA root cause
Other
Action
The firm is visiting consignees and making corrections as complaints are received. These corrections, to date, have been made between 04/04 and 09/06. The firm will correct other units as complaints are received and issued a Device Correction Letter dated 6/6 /07 notifying all users of the problem and appropriate responses should it occur.
Quantity in commerce
525 total for both models
Distribution
Worldwide: USA, Belgium, Canada, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Korea, Malaysia, Singapore and United Arab Emirates.
Product codes and lots
Serial numbers 0031 through 0324 are potentially affected by the problem. Serial numbers 0092, 0128, 0152, 0223, 0226, 0237, 0246, 0260 and 0296 have been corrected.