Z-0311-2007

Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required)...

Terumo Cardiovascular Systems Corporation / initiated 2004-05-01 / Terminated

Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763 on . The reason FDA records is "The battery status light on the front panel may not accurately reflect the actual status of the battery. The battery may contain more charge than indicated". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0311-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
The battery status light on the front panel may not accurately reflect the actual status of the battery. The battery may contain more charge than indicated.
FDA root cause
Other
Action
U.S. consignees were visited and new software installed between May 2004 and November 2005. Currently, approximately 50% of exported units have been upgraded.
Quantity in commerce
159
Distribution
Worldwide, including USA, Belgium, Canada, China, Egypt, Germany, India, Indonesia, Japan, Malaysia, Pakistan, Republic of China, Singapore, South Africa, South Korea and United Arab Emirates.
Product codes and lots
Serial numbers 0011 through 0039 and 0041 through 0170.