Z-0051-05
CIDEX OPA Solution, ortho-Phthalaldehyde Solution
Advanced Sterilization Products / initiated 2004-04-12 / Terminated
Advanced Sterilization Products began a recall of CIDEX OPA Solution, ortho-Phthalaldehyde Solution on . The reason FDA records is "Anaphylactic-like reactions were reported following repeated cystoscopy for bladder cancer patients where the scope had been reprocessed in CIDEX OPA Solution". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0051-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Advanced Sterilization Products
- Product
- CIDEX OPA Solution, ortho-Phthalaldehyde Solution
- Product code
- MED Sterilant, Medical Devices
- Reason for recall
- Anaphylactic-like reactions were reported following repeated cystoscopy for bladder cancer patients where the scope had been reprocessed in CIDEX OPA Solution.
- FDA root cause
- Other
- Action
- Firm conducted WebEx meeting with affiliates on April 12 & 13, 2004 to discuss CIDEX OPA Solution labeling. April 23, 2004, Firm sent letters to affiliates, customers, and end users to notify of new contraindication for CIDEX OPA Solution. Same letter was sent to U.S. Healthcare facilities on April 23, 2004; to Urology M.D.s on April 28, 2004; to Urologist Nurses on May 7, 2004. Customers can call 800-370-4632 with questions.
- Distribution
- Worldwide
- Product codes and lots
- Product Code: 20390, 20391, 20394