Z-1069-04
NxStage Cartridge Express. A disposable extracorporeal blood and fluid pathway and a hemo filter. Model...
NxStage Medical Inc / initiated 2004-04-14 / Terminated
NxStage Medical, Inc. began a recall of NxStage Cartridge Express. A disposable extracorporeal blood and fluid pathway and a hemo filter. Model #CAR-124, 6 units per case on . The reason FDA records is "Pinhole leaks on the replacement (RF) side FMP bag in certain lots of its disposable cartridge balance chambers resulting in loss of dialysate or ultrafiltrate". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1069-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- NxStage Medical Inc
- Product
- NxStage Cartridge Express. A disposable extracorporeal blood and fluid pathway and a hemo filter. Model #CAR-124, 6 units per case.
- Product code
- KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Reason for recall
- Pinhole leaks on the replacement (RF) side FMP bag in certain lots of its disposable cartridge balance chambers resulting in loss of dialysate or ultrafiltrate
- FDA root cause
- Other
- Action
- Voluntary recall notice dated 4/14/04 sent UPS, overnight with signature required to both ICU/CCU (Acute accounts) and Chronic accounts. Firm to remove and replace product that contains a disposable component from a new supplier that has demonstrated to be less reliable resulting in pinhole leaks.
- Quantity in commerce
- 234 units
- Distribution
- CA, IL, IN, MD, IA, MO, NJ, OH, VT
- Product codes and lots
- Lot Numbers from and including FG-031216-01 to FG-040322-01