Z-0988-04

Compounding Brochure titled ''Nutrition Compounding Simplified'', Lit #800415, print date January 2002; the...

Baxter Healthcare Corporation / initiated 2004-04-30 / Terminated

Baxter Healthcare Corp. began a recall of Compounding Brochure titled ''Nutrition Compounding Simplified'', Lit #800415, print date January 2002; the brochure is applicable to the Automix Plus High... on . The reason FDA records is "The promotional material for the Automix/Accusource and Micromix compounders and Micromix Compounder Implementation Guide contain an incomplete accuracy statement regarding the delivery of each ingredient by the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0988-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Compounding Brochure titled ''Nutrition Compounding Simplified'', Lit #800415, print date January 2002; the brochure is applicable to the Automix Plus High Speed Compounder System, product 2M8075, the Automix High Speed Compounder System, product 2M8077, the Automix 3+3 Compounder System, product 2M8286, the Automix 3+3 Compounder with Accusource Monitoring System, product 2M8287, the Micromix Compounder System, 1.0 mL, product 2M8290, and the Micromix Compounder System, 0.3 mL, product 2M8290L; Baxter Healthcare Corporation, Clintec Nutrition, One Baxter Parkway, Deerfield, IL 60015
Product code
LHI--
Reason for recall
The promotional material for the Automix/Accusource and Micromix compounders and Micromix Compounder Implementation Guide contain an incomplete accuracy statement regarding the delivery of each ingredient by the compounders.
FDA root cause
Other
Action
Baxter sent Important Product Information letters dated 4/30/04 to all of the compounder customers, informing them of the inaccurate information and requesting that the Nutrition Compounding Simplified brochure, Baxter Info Quick Cards on CD-Rom, versions Q4 2003 and Q2 2003, and pages 60 and 61 of the Micromix Compounder Implementation Guide be destroyed. Replacement pages 60 and 61 of the implementation guide were enclosed with the letter. The letters also listed the accuracy information for each compounder based on Baxter''s supporting data, as well as the specific information regarding the overdelivery alarm limits referenced in the Operator''s Manuals.
Quantity in commerce
1,366 units
Distribution
Nationwide, United Kingdom, Germany, Chile, Canada, Brazil, El Salvador, Taiwan, Korea, Italy, Saudi Arabia, Israel, China, Indonesia, Hong Kong, Colombia, Argentina, Mexico, Guatemala, Costa Rica, Panama and Honduras.
Product codes and lots
Lit #800415, print date January 2002  This brochure was distributed individually and as part of the Baxter Info Quick Card (on CD-Rom), versions Q4 2003 and Q2 2003