Biotronik Inc began a recall of Tachos ATx on . The reason FDA records is "Potential for device to not deliver full amount of energy". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0943-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- BIOTRONIK AG
- Product
- Tachos ATx
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- Potential for device to not deliver full amount of energy
- FDA root cause
- Other
- Action
- On 4/29/04 the firm issued 'Dear Doctor' letters to the physicians. The letter instructs the physician to schedule an immediate follow-up with all affected patients to determine if their devices are being impacted by the passivation buildup and to determine if device replacement is indicated.
- Quantity in commerce
- 50
- Distribution
- Devices were distributed to physicians and medical centers throughout the U.S. The firm does not distribute internationally.
- Product codes and lots
- Model Number 334 342 Serial numbers: 78050083 ¿ 78050088, 78050091 ¿ 78050095, 78050097, 78050098, 78050100, 78050103 ¿ 78050105, 78050107, 78050109, 78050110, 78050113, 78050117 ¿ 78050122, 78050124 ¿ 78050128, 78050133 ¿ 78050139, 78050142 ¿ 78050144, 78050162 ¿ 78050169