Z-0943-04

Tachos ATx

BIOTRONIK AG / initiated 2004-04-29 / Terminated

Biotronik Inc began a recall of Tachos ATx on . The reason FDA records is "Potential for device to not deliver full amount of energy". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0943-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BIOTRONIK AG
Product
Tachos ATx
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Potential for device to not deliver full amount of energy
FDA root cause
Other
Action
On 4/29/04 the firm issued 'Dear Doctor' letters to the physicians. The letter instructs the physician to schedule an immediate follow-up with all affected patients to determine if their devices are being impacted by the passivation buildup and to determine if device replacement is indicated.
Quantity in commerce
50
Distribution
Devices were distributed to physicians and medical centers throughout the U.S. The firm does not distribute internationally.
Product codes and lots
Model Number 334 342 Serial numbers: 78050083 ¿ 78050088, 78050091 ¿ 78050095, 78050097, 78050098, 78050100, 78050103 ¿ 78050105, 78050107, 78050109, 78050110, 78050113, 78050117 ¿ 78050122, 78050124 ¿ 78050128, 78050133 ¿ 78050139, 78050142 ¿ 78050144, 78050162 ¿ 78050169