Z-0919-04
VITEK GPS-107 Gram Positive Susceptibility Card, Catalog #V4368, for in-vitro diagnostic use, 20 cards per...
bioMerieux, Inc. / initiated 2004-04-20 / Terminated
bioMerieux Inc began a recall of VITEK GPS-107 Gram Positive Susceptibility Card, Catalog #V4368, for in-vitro diagnostic use, 20 cards per package. The responsible firm on the label is... on . The reason FDA records is "Some of the cards were stamped with an incorrect card code causing the system to read and report the cards incorrectly". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0919-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- VITEK GPS-107 Gram Positive Susceptibility Card, Catalog #V4368, for in-vitro diagnostic use, 20 cards per package. The responsible firm on the label is bioMerieux, Inc., Durham, NC.
- Product code
- LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
- Reason for recall
- Some of the cards were stamped with an incorrect card code causing the system to read and report the cards incorrectly.
- FDA root cause
- Other
- Action
- The firm issued recall letters dated 4/20/04 via certified mail requesting remaining inventories be destroyed and explaining the method which the customer can use to confirm the cards were not misreported.
- Quantity in commerce
- 1,760/20-card boxes
- Distribution
- Distribution was nationwide to medical facilities and laboratories. Foreign distribution was made to China and Switzerland.
- Product codes and lots
- Lot #P108X