Z-0811-04
Medtronic Micro Jewel II Implantable Cardioverter Defibrillators, Model 7223Cx
Medtronic, Inc. / initiated 2004-04-02 / Terminated
Medtronic Inc. Cardiac Rhythm Managment began a recall of Medtronic Micro Jewel II Implantable Cardioverter Defibrillators, Model 7223Cx on . The reason FDA records is "Some of the devices with suspect capacitors (high voltage capacitors associated with prior notifications in 1999 and in 2000) have had unexpected charge circuit time-outs or charge circuit inactive conditions as the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0811-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic Micro Jewel II Implantable Cardioverter Defibrillators, Model 7223Cx
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- Some of the devices with suspect capacitors (high voltage capacitors associated with prior notifications in 1999 and in 2000) have had unexpected charge circuit time-outs or charge circuit inactive conditions as the battery voltage nears the level for Elective Replacement of the devices.
- FDA root cause
- Other
- Action
- A letter, dated April 2004, was sent on April 5, 2004 to physicians following patients who had the defibrillators implanted. The letter gave recommendations to physicians for selective replacement of some of the implanted devices and for avoiding the problem. Physicians are being visited by sales representatives to verify notification.
- Quantity in commerce
- 5839 units (2394 units in the United States)
- Distribution
- Nationwide in the United States and worldwide.
- Product codes and lots
- Manufacturing dates from November 1996 to December 1997; and Use Before Dates from May 1998 to June 1999.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
