Z-0938-04

LIFEPAK 12 defibrillator/monitor

Physio Control, Inc. / initiated 2004-04-05 / Terminated

Medtronic Physio Control Corp began a recall of LIFEPAK 12 defibrillator/monitor on . The reason FDA records is "Potential for display to go blank due to an electrical component that may fail due to mechanical shock of vibration of the defibrillator". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0938-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Physio Control, Inc.
Product
LIFEPAK 12 defibrillator/monitor.
Product code
LDD Dc-Defibrillator, Low-Energy, (Including Paddles)
Reason for recall
Potential for display to go blank due to an electrical component that may fail due to mechanical shock of vibration of the defibrillator.
FDA root cause
Other
Action
Letters dated April 2004 were sent to customers on 4/6/04. The letter explains that a service representative will visit the customer and check the capacitor.
Quantity in commerce
2885
Distribution
Devices were distributed to hospitals and medical centers throughout the US. The firm also distributed units internationally.
Product codes and lots
Part numbers VLP12, UVLP12, EVLP12, DVLP12, AVLP12, and any of these may have five to eight digits after the base number.   Serial numbers can be obtained from the firm or FDA''s Seattle District Recall Coordinator