Z-0349-2027Class II
Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000...
Cochlear Limited / initiated 2026-09-30 / Open, Classified
Recall record
- Recall number
- Z-0349-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Cochlear Limited
- Product
- Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.
- Product code
- MCM Implant, Cochlear
- Reason for recall
- Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits
- FDA root cause
- Under Investigation by firm
- Action
- On 09/30/2026, an "URGENT Field Safety Corrective Action Notification" was sent to U.S. customers informing them that a potential condition has been identified that could result in one or more open circuits affecting individual electrode pathways with the implant. Customers were instucted to not implant and return affected products to Cochlear Limited. For questions contact [email protected].
- Quantity in commerce
- 4 device in the U.S.
- Distribution
- U.S.: NJ, NY, and PA. O.U.S.: Not provided
- Product codes and lots
- Model: CI1000 Product/Catalog: P1979576 UDI-DI code: 9321502069847 Serial Numbers: 1020210051972 1020210070461 1020210145046 1020210048286
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0349-2027 | Class II | Ongoing | Voluntary: Firm initiated |