Z-0333-2027Class II

Mobile column TruSystem 7500 U, Product Code 1730720

Baxter Healthcare Corporation / initiated 2026-09-14 / Open, Classified

Recall record

Recall number
Z-0333-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
Mobile column TruSystem 7500 U, Product Code 1730720
Product code
GDC Table, Operating-Room, Electrical
Reason for recall
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
FDA root cause
Equipment maintenance
Action
Customers were notified by telephone beginning 9/14/26. Baxter Healthcare Corporation is issuing a voluntary correction for the TruSystem Surgical Tables after identifying that four units underwent spindle replacement without the required belt tension verification. The affected units were released without confirmation that belt tension met the specified requirements, which could lead to unintended downward movement of the table column. Unintended downward movement of the surgical table could occur, potentially resulting in disruption of patient positioning, reduced perfusion or oxygenation, musculoskeletal injury and/or interruption of a surgical procedure. To date, Baxter has not received any reports of patient injury related to this issue. Baxter will call to schedule a time to come in and perform the belt tension verification. Customers can continue to use the tables.
Quantity in commerce
1 unit
Distribution
US Nationwide distribution in the states of IL, MO, TX.
Product codes and lots
UDI-DI: 00887761974388; Serial number: 110183858

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0333-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0332-2027Class IIOngoingVoluntary: Firm initiated
Z-0333-2027Class IIOngoingVoluntary: Firm initiated