Z-0347-2027Class II

Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray...

Shimadzu Corporation Medical Systems Division / initiated 2025-09-30 / Open, Classified

Recall record

Recall number
Z-0347-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Shimadzu Corporation Medical Systems Division
Product
Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.
FDA root cause
Process control
Action
On September 30, 2026 Shimadzu Corp. issued a Urgent Medical Device Recall Notification to affected consignees via email. Shimadzu asked consignees to take the following actions; 1. If the bolts become loose, an abnormal noise will occur when moving the C-arm. Please check for abnormal noises during pre-study inspections. 2. If an abnormal noise is heard, please stop using the product and contact our service department. 3. Please complete the Medical Device Recall Customer Acknowledgement.
Quantity in commerce
118 unts.
Distribution
US Nationwide distribution in the states of SC, TX, FL, TN, CA, FL, NY.
Product codes and lots
Serial number/Shimadzu Medical Systems #/UDI: M1F0D3BF4001/ SMS20250167/ (01)04540217067145(11)250; M1F0CC4EB001/ SMS20250098/ (01)04540217067022(11)241; M1F0D27EA001/ SMS20250001/ (01)04540217067138(11)241; 41F0F43E7001/ SMS20250034/ (01)04540217068951(11)240; M1F0D3BE7001/ SMS20250248/ (01)04540217067145(11)240; M1F0D3BE6001/ SMS20240196/ (01)04540217067145(11)240; 41F0F40E5001/ N/A/ (01)04540217068937(11)240; 41F0F41E5001/ N/A/ (01)04540217068944(11)240; 41F0F43E5001/ SMS20240236/ (01)04540217068951(11)240; M1F0CD7E4001/ SMS20240162/ (01)04540217067046(11)240; 41F0F41E1001/ SMS20240062/ (01)04540217068944(11)240; 41F0F43DB002/ SMS20240101/ (01)04540217068951(11)231; 41F0F43DB001/ SMS20240100/ (01)04540217068951(11)231; M1F0CD7D4001/ SMS20230178/ (01)04540217067046(11)230

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0347-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0347-2027Class IIOngoingVoluntary: Firm initiated