Z-0359-2027Class II
LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303
Luminex Corporation / initiated 2026-10-05 / Open, Classified
Recall record
- Recall number
- Z-0359-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Luminex Corporation
- Product
- LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303
- Product code
- NSU Instrumentation For Clinical Multiplex Test Systems
- Reason for recall
- Instructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue may contribute to incorrect or delayed diagnosis of bloodstream infection.
- FDA root cause
- Error in labeling
- Action
- On October 5, 2026, the firm began notifying customers via "Urgent Field Safety Notice - Medical Device Recall" letters. Consignees were informed that the current Instructions for Use for the Gram-Negative Blood Culture Assay do not adequately describe interpretation of certain positive family-level results. Specifically, when a positive family-level only result is obtained ("family-level only result" refers to a positive organism family-level detection without an accompanying genus- or species-level identification), repeat testing with the LIAISON PLEX Gram-Negative Blood Culture Assay is recommended. Absence of this instruction may increase the likelihood that certain reportable detections are not appropriately recognized by the user and may contribute to an incorrect interpretation of test results. Customers were instructed to implement the following interpretation guidance: When a positive family-level only result is obtained, repeat testing with the LIAISON PLEX BCN assay is recommended. DiaSorin is updating the Instructions for Use to include the revised guidance. Revised labeling will be provided following implementation of the labeling revision. Please contact Technical Support Services with any questions or concerns at the following phone numbers:1-877-785-2323 (U.S. and Canada); +1-512-381-4397 (Outside U.S. and Canada); 1-512-219-5114 (Fax) and email:[email protected].
- Quantity in commerce
- 30,168 test cartridges (2,514 Kits)
- Distribution
- US distribution nationwide. No OUS distribution.
- Product codes and lots
- All lots UDI-DI pending
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0359-2027 | Class II | Ongoing | Voluntary: Firm initiated |