Z-0322-2027Class II

NovaSeq 6000Dx Instrument REF: 20068232, FOR IN VITRO DIAGNOSTIC USE

Illumina, Inc. / initiated 2024-05-08 / Open, Classified

Recall record

Recall number
Z-0322-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Illumina, Inc.
Product
NovaSeq 6000Dx Instrument REF: 20068232, FOR IN VITRO DIAGNOSTIC USE
Product code
PFF High Throughput Dna Sequence Analyzer
Reason for recall
(Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission.
FDA root cause
Labeling design
Action
On May 8, 2024, Illumina, Inc. issued a Urgent Medical Device Recall via email. Illumina ask consignees to take the following 1. Download & read this Cybersecurity Guidance document. The full hyperlink is below. https://support.illumina.com/support-content/cyberguidance.html 2. For Each Product Connected to the Customer Network: work with your internal IT department to determine which secure protocol works best in your environment and to implement the appropriate configuration change. Note: a change to the network protocol affects both Dx and RUO modes simultaneously therefore there is no need to boot into each mode separately to take the necessary action. 3. Complete and return the Verification Form after carrying out all the steps in the instructions provided on your specific product(s) identified as affected in Table 1.
Quantity in commerce
41 units
Distribution
Worldwide - US Nationwide distribution including in the states and territory of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of AE, AR, AT, AU, BE, BG, CA, CH, CL, CN, CY, CZ, DE, DK, DZ, EG, ES, FI, FR, GB, Ghana, GR, GT, HK, HU, IE, IL, IS, IT, JP, KR, KW, LT, LU, LV, MA, MT, NG, NL, NO, NZ, PE, PH, PL, PT, QA, RO, RU, SA, SE, SG, SI, TH, TR, UA, VN, ZA.
Product codes and lots
All Lots/UDI:00816270020637

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0322-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0320-2027Class IIOngoingVoluntary: Firm initiated
Z-0321-2027Class IIOngoingVoluntary: Firm initiated
Z-0322-2027Class IIOngoingVoluntary: Firm initiated
Z-0323-2027Class IIOngoingVoluntary: Firm initiated
Z-0324-2027Class IIOngoingVoluntary: Firm initiated