Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0180-2027Class II

Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660...

Medline Industries, LP / initiated 2026-08-31 / Open, Classified

Recall record

Recall number
Z-0180-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medline Industries, LP
Product
Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ908720A
Product code
FDE Laparoscopy Kit
Reason for recall
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
FDA root cause
Process change control
Action
The firm began notifying customers on August 31, 2026. Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays. Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device. Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use. The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu.
Quantity in commerce
3,551 kits
Distribution
US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
Product codes and lots
DYNJ911135 UDI-DI 10198459264382 Lots 25EDA279 25FDB612 25LDB658 DYNJ911660 UDI-DI 10198459487071 Lots 25IMF444 25JMA458 26BMG519 26DMA914 26EMA779 26FME775 26GMA849 26GMF826 26GMM101 26GMM127 DYNJ911560 UDI-DI 10198459400261 Lots 25HMG749 25JMC886 25KMA383 25KMA392 DYNJ908720A UDI-DI 10195327423155 Lots 24DBN292 24GBJ114 24IBK504 24JBC471 25KBP733 26ABG732

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0180-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0180-2027Class IIOngoingVoluntary: Firm initiated
Z-0181-2027Class IIOngoingVoluntary: Firm initiated
Z-0182-2027Class IIOngoingVoluntary: Firm initiated
Z-0183-2027Class IIOngoingVoluntary: Firm initiated
Z-0184-2027Class IIOngoingVoluntary: Firm initiated
Z-0185-2027Class IIOngoingVoluntary: Firm initiated