Z-0167-2027Class II

TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519

Howmedica Osteonics Corp. / initiated 2026-09-16 / Open, Classified

Recall record

Recall number
Z-0167-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Howmedica Osteonics Corp.
Product
TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.
Product code
FSM Tray, Surgical, Instrument
Reason for recall
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
FDA root cause
Labeling design
Action
Stryker notified consignees on about 09/16/2026 via "MEDICAL DEVICE CORRECTION" letter mailed through UPS 2-day air. Consignees were informed that reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes, however, these trays require a sterilization dry time of 40 minutes when sterilized in blue wrap. Consignees were instructed to inform users of the Medical Device Correction and forward the notice to all individuals who need to be made aware or organizations who have consigned product, review and follow the updated instructions provided in LSTPI-B Cleaning, Sterilization, Inspection, and Maintenance of Reusable Medical Devices Reference Guide, complete and return the provided Medical Device Correction Business Reply Form.
Quantity in commerce
577 trays
Distribution
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Product codes and lots
Catalog Number: 6543-6-519. GTIN: 07613327555127. All lot numbers used prior to the correction date of May 8, 2026 are impacted.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0167-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0164-2027Class IIOngoingVoluntary: Firm initiated
Z-0165-2027Class IIOngoingVoluntary: Firm initiated
Z-0166-2027Class IIOngoingVoluntary: Firm initiated
Z-0167-2027Class IIOngoingVoluntary: Firm initiated