Z-0107-2027Class II
SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000
Polymedco, Inc. / Polymedco Cancer Diagnostic Products, Llc / initiated 2026-08-25 / Open, Classified
Recall record
- Recall number
- Z-0107-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Polymedco, Inc. / Polymedco Cancer Diagnostic Products, Llc
- Product
- SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000;
- Product code
- JPH Test, Erythrocyte Sedimentation Rate
- Reason for recall
- During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).
- FDA root cause
- Process control
- Action
- On August 26, 2026, URGENT PRODUCT RECALL letters were sent to customers. ACTIONS REQUIRED: 1. Examine your inventory and identify all Sediplast Westergren ESR tubes from Lot E1150A. 2. Immediately discontinue use and quarantine all unused tubes from the affected lot. 3. If affected product was distributed to other facilities or locations, immediately forward this notification to those recipients. 4. Contact Polymedco Technical Services to arrange replacement of the affected product and obtain instructions for its return or disposal. 5. Complete and return the enclosed Customer Recall Response Form to Polymedco within five business days, even if you no longer have any affected product in inventory. 6. If affected tubes were previously used without an incident, no additional action is required. If a cap-ejection, blood-spray or exposure incident occurred, follow your facility s applicable biohazard-exposure procedures and report the incident to Polymedco Technical Services. Please do not discard quarantined product until you have received return or disposal instructions from Polymedco.
- Quantity in commerce
- 448400
- Distribution
- Worldwide - US Nationwide distribution in the states and territory of AK, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV and the country of United Kingdom.
- Product codes and lots
- Model/Catalog No. S-100 and S-1000; UDI-DI: N/A; Lot Number E1150A; Expiration Date 06/30/27
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0107-2027 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY