Z-0087-2027Class II
Rollator, Bariatric, Basic Steel, Blue, mechanical walker with four legs and a metal frame intended for...
Medline Industries, LP / initiated 2026-08-14 / Open, Classified
Recall record
- Recall number
- Z-0087-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medline Industries, LP
- Product
- Rollator, Bariatric, Basic Steel, Blue, mechanical walker with four legs and a metal frame intended for medical purposes to provide moderate weight support while walking, Product Number MDS86800XWBS
- Product code
- ITJ Walker, Mechanical
- Reason for recall
- The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from the rollator frame tubing, the front wheels may become loose or detach from the rollator. If the front wheels become loose or detach during use, there is risk of severe adverse events. The immediate consequences could include falls resulting in head trauma, bone fractures, or other critical injuries.
- FDA root cause
- Process control
- Action
- An URGENT MEDICAL DEVICE RECALL notification letter was sent to customers beginning 8/14/26. REQUIRED ACTION: 1. Inspect Your Rollator Immediately " Please refer to Image 1: How to locate and identify the product Lot Number to locate the front and rear side of the rollator. " On the rear side of the rollator, you will find the Lot Label is located on the left side frame tubing, below the seat frame tubing. 2. If your rollator matches both the affected SKU and an affected Lot Number, immediately discontinue use of the device. Take a clear photograph of the Lot Label and contact Medline Industries, LP directly, as soon as possible. A photograph of the Lot Number must be submitted to Medline for verification. You may contact Medline Industries, LP using one of the following methods: " Email: [email protected] " Phone: 1-800-950-0128 option 5 Once Medline receives and confirms your Lot Number, further instructions regarding discontinued use of the affected rollator will be provided. 3. If your rollator does not have one of the affected Lot Numbers, it may continue to be used in accordance with the Instructions for Use. Please complete and return the attached response form only if you do not have the affected rollator Lot Number. 4. If you have resold, distributed, or transferred this product to another company or individual, FDA regulations require that you notify those recipients of this recall communication. Please contact Medline Industries, LP for further instructions regarding the identification, notification, and handling of affected products distributed to your customers. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
- Quantity in commerce
- 835 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Product codes and lots
- Lot Numbers: 05425080001 05426040001
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0086-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0087-2027 | Class II | Ongoing | Voluntary: Firm initiated |