Z-0207-2027Class II

Techno-Path Multichem AE Unassayed Level 3, REF:AE623X

Techno-Path Manufacturing Ltd. / initiated 2026-08-14 / Open, Classified

Recall record

Recall number
Z-0207-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Techno-Path Manufacturing Ltd.
Product
Techno-Path Multichem AE Unassayed Level 3, REF:AE623X
Product code
OHQ Multi-Analyte Controls Unassayed
Reason for recall
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
FDA root cause
Under Investigation by firm
Action
On August 14, 2026, Techno Path Manufacturing Ltd has issued a medical device correction to consignees via email. Techno Path Manufacturing Ltd informed consignees to please verify their inventory against the affected product codes and lot numbers. Techno Path Manufacturing Ltd. has informed the consignees to take the following actions: 1. Immediately place all affected inventory on hold and prevent further distribution. 2. Quarantine all affected product currently under your control. 3. Confirm that there has been no distribution of the impacted lots to customers. 4. Confirm receipt of this notification and provide inventory and distribution details to Technopath Manufacturing Ltd. within 3 business days. 5. Arrange return or destruction of affected inventory in accordance with instructions provided by Technopath Manufacturing Ltd.
Quantity in commerce
832 units
Distribution
US Nationwide distribution in the state of Massachusetts.
Product codes and lots
UDI:(01)05391523442824/-Lot # 65705263

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0207-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0205-2027Class IIOngoingVoluntary: Firm initiated
Z-0206-2027Class IIOngoingVoluntary: Firm initiated
Z-0207-2027Class IIOngoingVoluntary: Firm initiated