Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0134-2027Class II

JenaValve Trilogy Delivery Catheter, Size 23MM REF: TRILOGY-DC-S

Jenavalve Technology Inc. / initiated 2026-07-31 / Open, Classified

Recall record

Recall number
Z-0134-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Jenavalve Technology Inc.
Product
JenaValve Trilogy Delivery Catheter, Size 23MM REF: TRILOGY-DC-S
Product code
NPT Aortic Valve, Prosthesis, Percutaneously Delivered
Reason for recall
Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.
FDA root cause
Nonconforming Material/Component
Action
On the dates of August 1, 4, 6, and 19th, JenaValve Technology Inc. issued a Urgent: Medical Device Recall notification via email. JenaValve asked consignees to take the following actions: 1. Review the information contained in this letter and ensure that anyone in your facility who needs to be aware of this notification reads this letter carefully. 2. Examine your inventory immediately to determine if you have Trilogy Delivery Catheters from the affected lots and quarantine the subject products. 3. JenaValve field team members will schedule a visit to retrieve the subject Trilogy Delivery Catheters. If you prefer to return the products yourself, please contact us at 866-588-8287 and we will provide you with an RGA number prior to you shipping them to: JenaValve Technology Inc., 4 Cromwell, Irvine, California 92618, USA; Attn: Quality Department; Reference: FA26001 4. Please complete and sign the Business reply form (BRF) of this letter and email it to [email protected] or give it to your JenaValve field representative. 5. Upon receipt of the completed BRF and returned product, JenaValve will provide replacement products. Incomplete BRFs cannot be processed. 6. If any of the subject Trilogy Delivery Catheters have been transferred to another facility, identify the receiving facility on the Business Reply Form, contact [email protected], and provide a copy of this letter to the relevant personnel.
Quantity in commerce
111 units
Distribution
US Nationwide distribution in the states of AZ, CA, FL, GA, IL, MA, MD, MI, MN, MO, NY, OH, PA, SC, TX, UT, VA, WA and the countries of Germany, Denmark, Great Britain, Italy.
Product codes and lots
GTIN - 00850000898134/ Lot: DS-040826-003 DS-042326-003, DS-042726-006, DS-042826-005, DS-050426-007, DS-050626-002, DS-050626-006, DS-050626-009, DS-050726-005

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0134-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0134-2027Class IIOngoingVoluntary: Firm initiated
Z-0135-2027Class IIOngoingVoluntary: Firm initiated
Z-0136-2027Class IIOngoingVoluntary: Firm initiated