Z-1875-2013Class II

Mobiletta Mira Mobile X-ray System

Siemens Medical Solutions USA, Inc. / initiated 2013-05-02 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Mobiletta Mira Mobile X-ray System on . FDA classified it as Class II and gives the reason as "Siemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1875-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Mobiletta Mira Mobile X-ray System
Product code
IZL System, X-Ray, Mobile
Reason for recall
Siemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or elevators), it is possible that one of its main wheel shafts could break because of material fatigue, resulting in the chassis falling on the ground on the affected wheel side. This
Root cause tag
design
FDA root cause
Device Design
Action
Siemens sent a Field Safety Notice dated May 2. 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed to : 1) Never step or sit on the unit or otherwise put substantial weight on it. 2) Avoid moving the unit over high thresholds for doors or elevators where possible, or do so carefully and slowly. 3) Make sure that no person had his/her extremities under the chassis, also when the unit is not moving. Update instructions were also released to alert customers of the potential risk and to provide a hardware change which is intended to improve corrosion resistance. An acknowledgement of receipt form was also sent out for the field safety notice and requested to be returned to Siemens. For further questions please call (888) 826-9702.
Quantity in commerce
39
Distribution
US Distribution including the states of IA, SD, NC, WI, OH, NE, TX, KS, ID, MI, GA, AR and IL.
Product codes and lots
Model number 10273100

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1875-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1875-2013Class IITerminatedVoluntary: Firm initiated