Mobiletta Mira Mobile X-ray System
Siemens Medical Solutions USA, Inc began a recall of Mobiletta Mira Mobile X-ray System on . FDA classified it as Class II and gives the reason as "Siemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1875-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Mobiletta Mira Mobile X-ray System
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- Siemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or elevators), it is possible that one of its main wheel shafts could break because of material fatigue, resulting in the chassis falling on the ground on the affected wheel side. This
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens sent a Field Safety Notice dated May 2. 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed to : 1) Never step or sit on the unit or otherwise put substantial weight on it. 2) Avoid moving the unit over high thresholds for doors or elevators where possible, or do so carefully and slowly. 3) Make sure that no person had his/her extremities under the chassis, also when the unit is not moving. Update instructions were also released to alert customers of the potential risk and to provide a hardware change which is intended to improve corrosion resistance. An acknowledgement of receipt form was also sent out for the field safety notice and requested to be returned to Siemens. For further questions please call (888) 826-9702.
- Quantity in commerce
- 39
- Distribution
- US Distribution including the states of IA, SD, NC, WI, OH, NE, TX, KS, ID, MI, GA, AR and IL.
- Product codes and lots
- Model number 10273100
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1875-2013 | Class II | Terminated | Voluntary: Firm initiated |