Z-1923-2013Class II
BD Vacutainer Multiple Sample and Direct Draw Luer Adapter BD Franklin Lakes, NJ 07417 USA A sterile...
Becton Dickinson & Co / initiated 2013-05-01 / Terminated
Becton Dickinson & Company began a recall of BD Vacutainer Multiple Sample and Direct Draw Luer Adapter BD Franklin Lakes, NJ 07417 USA A sterile non-invasive device used to connect venous access devices... on . FDA classified it as Class II and gives the reason as "BD Vacutainer Multiple Sample and Direct Draw Luer Adapter may have blood leakage and low draw volume". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1923-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Vacutainer Multiple Sample and Direct Draw Luer Adapter BD Franklin Lakes, NJ 07417 USA A sterile non-invasive device used to connect venous access devices such as needles, blood collection sets, ad infusion sets to blood collection tubes.
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- BD Vacutainer Multiple Sample and Direct Draw Luer Adapter may have blood leakage and low draw volume.
- FDA root cause
- Under Investigation by firm
- Action
- BD sent a Product Advisory Notice dated May 3, 2013, via UPS to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discard filled tubes since it is possible they will contain less blood volume (low draws), replace the Luer Adaptor unit being used with a fresh unit, and continue to draw blood samples using the fresh unit. Customers with questions were instructed to contact BD Technical Service at 1-800-631-0174, Option 4, or contact them through website: www.bd.com/vacutainer/contact. For questions regarding this recall call 201-847-6800.
- Quantity in commerce
- 37,906,400 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Marketing Status: Class II, 510K # K991088 Device Listing Number: D030347 Catalog number 367290, 367300 (CE), 368491, and 303380.