Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1525-2013Class II

Siemens brand MicroScan Dried Negative Breakpoint Combo 47 panels (B1017-417), SMN #: 10483099), For use in...

Siemens Healthcare Diagnostics Inc. / initiated 2013-05-02 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens brand MicroScan Dried Negative Breakpoint Combo 47 panels (B1017-417), SMN #: 10483099), For use in determining quantitative and/or qualitative... on . FDA classified it as Class II and gives the reason as "The product has the potential for false negative or delayed reaction for Arginine (ARG) results. In addition, 42 hour identification panel holds may occur due to a weak ARG reaction". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1525-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens brand MicroScan Dried Negative Breakpoint Combo 47 panels (B1017-417), SMN #: 10483099), For use in determining quantitative and/or qualitative antimicrobial agent susceptibility.
Product code
LTT Panels, Test, Susceptibility, Antimicrobial
Reason for recall
The product has the potential for false negative or delayed reaction for Arginine (ARG) results. In addition, 42 hour identification panel holds may occur due to a weak ARG reaction.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens sent an Urgent Field Safety Notice dated May 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact your local Siemens Healthcare Diagnostics Representative for information concerning this notice.
Quantity in commerce
683 units
Distribution
Nationwide Distribution.
Product codes and lots
Lot: 2013-11-13, Exp 11/13/2013.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1525-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1525-2013Class IITerminatedVoluntary: Firm initiated