Z-1402-2013Class II

Integrity 3.0

Elekta Limited / initiated 2013-04-30 / Terminated / root cause: manufacturing process

Elekta, Inc. began a recall of Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to... on . FDA classified it as Class II and gives the reason as "Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1402-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Elekta sent an Important Field Safety Notice (IFSN) letter dated April 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers are instructed to complete and return the FCO Action Notification Report to local Elekta Office or Representative.
Quantity in commerce
31
Distribution
USA Nationwide including the states of CA, FL, IL, MD, MA, NJ, NY, NC, ND, OH, OR, PA, TN, TX, VA, WA, DC and Puerto Rico
Product codes and lots
325134, 325254, 325082, 325159, 325268, 325237, 325331, 325267, 325236, 325314, 325127, 325261, 325286, 325077, 325269, 325129, 325296, 325273, 325139, 325201, 325137, 325184, 325161, 325076, 325086, 325112, 325081, 325083, 325099, 325110, 325271

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1402-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1402-2013Class IITerminatedVoluntary: Firm initiated