Integrity 3.0
Elekta, Inc. began a recall of Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to... on . FDA classified it as Class II and gives the reason as "Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1402-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Elekta sent an Important Field Safety Notice (IFSN) letter dated April 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers are instructed to complete and return the FCO Action Notification Report to local Elekta Office or Representative.
- Quantity in commerce
- 31
- Distribution
- USA Nationwide including the states of CA, FL, IL, MD, MA, NJ, NY, NC, ND, OH, OR, PA, TN, TX, VA, WA, DC and Puerto Rico
- Product codes and lots
- 325134, 325254, 325082, 325159, 325268, 325237, 325331, 325267, 325236, 325314, 325127, 325261, 325286, 325077, 325269, 325129, 325296, 325273, 325139, 325201, 325137, 325184, 325161, 325076, 325086, 325112, 325081, 325083, 325099, 325110, 325271
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1402-2013 | Class II | Terminated | Voluntary: Firm initiated |